257
error
-
File name
256
not found
107
File
transfer
failed
56
Fixture
Simulator
not
allowed
24
Frequency
offset
enabled
59
Frequency
offset not
allowed
25
Frequency
offset not
enabled
62
Frequency
out of
range
G
No.
Message
-105
GET not
command is not executed correctly. This error
occurs when you try to copy to an unsuitable file
name, for example.
The file name specified is not found and hence the
command is not executed correctly. This error
occurs when you try to read a file that does not
exist in a disk or a disk is not correctly inserted
into the drive to read or write a file, for example.
This error occurs when writing data into or
reading data from a file (MMEM:TRAN command)
fails.
This error occurs when the fixture simulator is
executed (CALC:FSIM:STAT ON) under the
condition of the frequency offset is enabled
(SENS:OFFS ON).
The fixture simulator is allowed to use when the
frequency offset is not enabled.
This error occurs when the normal calibration
(SENS:CORR:COLL:OPEN etc.) is performed
under the condition of the frequency offset is
enabled (SENS:OFFS ON).
This error occurs when frequency offset function
is not allowed in the current setting.
This error occurs when the scalar-mixer
calibration (SENS:CORR:OFFS:COLL:OPEN etc.) is
performed under the condition of the frequency
offset is not enabled (SENS:OFFS OFF).
This error occurs when the setting of frequency
exceeds output frequency range under the
condition of the frequency offset is enabled
(SENS:OFFS ON).
Description
A group execution trigger (GET) has been
received in the program message (see 7.7,
Product Information
761