Agilent Technologies E5071C Manual page 189

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

E5071C
3. Click Modify Kit. To change the pre-defined connector type (e.g.
OPEN(f) to OPEN (m)) skip to the
4. Click Define STDs.
5. Select the standard to be redefined from among standards numbered
1 through 21.
6. Click No Name > Label and input a new label for the standard using
the keypad displayed on the screen.
Procedure to select the standard type and define standard coefficient
1. Press
key.
Cal
2. Click Modify Kit.
3. Click Define STDs and select the standard to be redefined from
among standards numbered 1 through 21.
4. Click STD Type, then select the
Softkey
Open
Short
Load
Delay/Thru
Unknown Thru
Arbitrary
None
5.
6. Set the
standard
Softkey
C0
188
specifty
type of
Function
Selects the OPEN standard
Selects the SHORT standard
Selects the LOAD standard
Selects the delay/THRU standard
Selects the Unknown THRU standard
Selects the arbitrary impedance
Selects no standard type
coefficient.
Function
Sets C0
classes.
standard.