Selecting Target Port Of Error Correction; Setting Loss Compensation; Turning On/Off Loss Compensation - Agilent Technologies E5071C Manual

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

2. Click Power Calibration > Sensor A/B Settings > Zero Type > INT|EXT
to select from internal or external zeroing.
3. Click Power Calibration > Sensor A/B Settings > Zero/Calibrate Sensor
4. For U2000 Series USB power sensors, Sensor B setting is
disabled.

Selecting Target Port of Error Correction

The power level error correction is executed for each channel/test port.
You can set the following items for each channel/test port:
ON/OFF of error correction
Setting of loss compensation
Selection of the power sensor
Number of power level measurements at one measurement point
Calibration data
Follow the steps below to select the test port for which you want to
set/execute power level error correction.
1. Press
Cal
2. Click Power Calibration.
3. Press
Channel Next/Channel Prev
4. Click Select Port.
5. Click the softkey corresponding to the test port you want to select.

Setting Loss Compensation

If you need to apply a signal with a more accurate power level to the DUT,
you can use the loss compensation feature to obtain calibration data for
correcting the difference in power loss due to the difference in connection
method (cable, adapter, etc.) between the time when the power calibration
data are measured and when the actual DUT is measured.
The loss compensation feature corrects the power measurement result
based on the preset power loss data. More specifically, if you turn on the
loss compensation, the measurement result of the power level obtained in
the power calibration data measurement is a value obtained by adding the
loss value set in the loss compensation table to the measured power value.

Turning ON/OFF loss compensation

Follow the steps below to turn ON/OFF the loss compensation.
1. Press
Cal
2. Click Power Calibration.
3. Select a port (refer to Selecting target port of error correction).
key.
key.
keys to select the channel.