Agilent Technologies E5071C Manual page 618

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

1. Turn on the E5071C.
2. Press
Macro Setup
3. Click Load Project... in the softkey menu.
4. Select RestoreSysCorFile.vba from the D:\Agilent\Service folder and
then press the Open button.
5. Press
Macro Run
6. The RestoreSysCorFile dialog box appears.
7. Click OK.
8. Once the program has terminated, restart the E5071C to enable
the system correction data.
key.
key