Agilent Technologies E5071C Manual page 141

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

E5071C
3. Click Calibrate > 2-Port Cal.
4. Click Select Ports, then select the test ports on which you will
perform full 2-port calibration. (In the procedure below, the selected
test ports are denoted as x and y.)
5. Click Reflection.
6. Connect an OPEN calibration standard to test port x (the connector
to which the DUT is to be connected).
7. Click Port x Open to start the calibration measurement (x denotes the
test port to which the standard is connected).
If you select the calibration kit which has different
o
calibration definitions for each gender, (m) and (f) in the
name (label) of the standard displayed in the softkey
indicate male (m) and female (f) for the analyzer's
connector, respectively.
8. Disconnect the OPEN calibration standard and replace it with a
SHORT calibration standard.
9. Click Port x Short to start the calibration measurement (x denotes the
test port to which the standard is connected).
10.
Disconnect the SHORT calibration standard and replace it with a
LOAD standard.
11.
Click Port x Load to start the calibration measurement (x denotes
the test port to which the standard is connected).
12.
Repeat the above procedure for port y.
13.
Click Return.
14.
Click Transmission.
15.
Make a THRU connection between ports x and y (between the
connectors to which the DUT is to be connected).
16.
Click Port x-y Thru to start the calibration measurement (x and y
denote the test ports between which the THRU connection is being
made).
17.
Click Return.
18.
If an isolation calibration must be performed using a LOAD
standard, follow the procedure below.
19.
Click Isolation (Optional).
20.
Connect a LOAD standard to each of the two test ports
(connectors to which the DUT is to be connected).
140