Measure:vpp - Agilent Technologies InfiniiVision 3000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 3000 X-Series:
Table of Contents

Advertisement

:MEASure:VPP

Command Syntax
:MEASure:VPP [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels) in NR1 format
<r> ::= 1-2 in NR1 format
The :MEASure:VPP command installs a screen measurement and starts a
vertical peak- to- peak measurement. If the optional source parameter is
specified, the current source is modified.
Query Syntax
:MEASure:VPP? [<source>]
The :MEASure:VPP? query measures the maximum and minimum vertical
value for the selected source, then calculates the vertical peak- to- peak
value and returns that value. The peak- to- peak value (Vpp) is calculated
with the following formula:
Vmax and Vmin are the vertical maximum and minimum values present on
the selected source.
Return Format
<value><NL>
<value> ::= vertical peak to peak value in NR3 format
See Also
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
(see
page
1138)
Vpp = Vmax - Vmin
"Introduction to :MEASure Commands"
":MEASure:SOURce"
":MEASure:VMAX"
on page 459
":MEASure:VMIN"
on page 460
":MEASure:VAMPlitude"
on page 406
on page 444
on page 456
20
:MEASure Commands
461