:CALibrate:OUTPut

Command Syntax
:CALibrate:OUTPut <signal>
<signal> ::= {TRIGgers | SOURce | DSOurce | MASK}
The CALibrate:OUTPut command sets the signal that is available on the
rear panel TRIG OUT BNC:
• TRIGgers — pulse when a trigger event occurs.
• SOURce — raw output of trigger comparator.
• DSOurce — SOURce frequency divided by 8.
• MASK — signal from mask test indicating a success or fail mask test.
Query Syntax
:CALibrate:OUTPut?
The :CALibrate:OUTPut query returns the current source of the TRIG OUT
BNC signal.
Return Format
<signal><NL>
<signal> ::= {TRIG | SOUR | DSO | MASK}
See Also
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
(see
page
658)
"Introduction to :CALibrate Commands"
":MTESt:OUTPut"
on page 336
Commands by Subsystem
on page 179