:CALibrate:OUTPut

Command Syntax
:CALibrate:OUTPut <signal>
<signal> ::= {TRIGgers | MASK | WAVEgen}
The CALibrate:OUTPut command sets the signal that is available on the
rear panel TRIG OUT BNC:
• TRIGgers — pulse when a trigger event occurs.
• MASK — signal from mask test indicating a failure.
• WAVEgen — waveform generator sync output signal. This signal depends
Waveform
Type
SINusoid,
SQUare,
RAMP,
PULSe
DC, NOISe
Query Syntax
:CALibrate:OUTPut?
The :CALibrate:OUTPut query returns the current source of the TRIG OUT
BNC signal.
Return Format
<signal><NL>
<signal> ::= {TRIG | MASK | WAVE}
See Also
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
(see
page
790)
on the :WGEN:FUNCtion setting:
Sync Signal Characteristics
The Sync signal is a TTL positive pulse that occurs when the waveform rises
above zero volts (or the DC offset value).