Impedance Measurement Accuracy Test - Agilent Technologies E4991A Service Manual

Rf impedance/material analyzer
Hide thumbs Also See for E4991A:
Table of Contents

Advertisement

Impedance Measurement Accuracy Test

This test verifies the E4991A's impedance measurement accuracy with a calibrated
standards(16190B).
Specification
Basic Accuracy:
NOTE
See the "Specifications and Supplemental Performance Characteristics" of the Operation
Manual.
Required Equipment
Performance Test Kit
Calibration Kit
Procedure
Preparation
1. Connect the equipment as shown in Figure 2-5.
Figure 2-5
Impedance Measurement Accuracy Test
2. Press
3. Press
4. Press
5. Set the measurement points using Add Segment as Figure 2-6.
Chapter 2
Description
to preset the E4991A.
[Preset]
.
[Sweep]
to show segment table.
Segment Table Menu
Performance Test and Function Test
0.65%
Recommended Model
16190B
16195B(furnished with E4991A)
Performance Test
35

Advertisement

Table of Contents
loading

Table of Contents