Fixture Simulator
Converting the Port Impedance of the Measurement Result
Converting the Port Impedance of the Measurement Result
The measured value obtained by using a port impedance of 50 Ω can be converted into a
measured value at an arbitrary port impedance.
Figure 7-3
Port impedance conversion function
Converting the Port Impedance
Step 1. Press
Step 2. Specify the port reference impedance for the required port.
•
•
Step 3. Press
state.
Step 4. Press
Step 5. If
286
-
Fixture Simulator - Port Z Conversion
Select
,
Port 1 Z0 Real
Port 2 Z0 Real
reference impedance in "Real" format.
Select
and
Port 1 Z0 Real
and
Z0 Real
Port 3 Z0 Imag
reference impedance in "Complex" format.
to change the port impedance conversion function to the
Port Z Conversion
.
Return
is
, press the key again to turn it
Fixture Simulator
OFF
.
,
Port 3 Z0 Real
,
Port 1 Z0 Imag
Port 2 Z0 Real
, or
and
Port 4 Z0 Real
Port 4 Z0 Imag
, or
to set the port
Port 4 Z0 Real
and
,
Port2 Z0 Imag
Port 3
to set the port
ON
.
ON
Chapter 7