HP 8722D Measurement Port Specifications

The following specifications show the residual system performance, including switch
repeatability, after error-correction. The error-correction consists of a fuII Z-port measurement
calibration, including isolation, with an IF bandwidth of 10 Hz and the specified calibration
kit. Calibration temperature is 23O f3OC. Measurement must be performed &lo of calibration
temperature.

HP 8722D with 2.4 mm Connectors

The following specifications describe the system performance of the HP 8722D network
analyzers. The system hardware includes the following:
Options: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Standard
Calibration kit: . . . . . . . . . , . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 85056A
Source match
Load match
Reflection tracking
Transmission tracking
I
0.01
-10
10
0.05
0.00
0.2
0.0
7-16
Specifications and Measurement Uncertainties
0.05 to 2 GHz
42 dB
41 dB
42 dB
-30
-50
-70
-90
0.4
0.6
0.8
1.0
Frequency Range
8 to 20 GHz
38 dB
38 dB
42 dB
42 dB
10.0
0.2
20 to 40 GEz
38 dB
33 dB
38 dB
-30
-50
0.4
0.6
0.8