To Set Averaging Rate|Stabilizing the measurement result

1. Press
.
2. Enter the averaging rate using the numeric ENTRY keys. (For example, to enter 4, press
.) You can enter integer values from 1 to 256. Also, you can increase or decrease the
value using
or
3. Press
to set the value and to exit.

To Select Trigger Mode

Press
until the Trigger annunciator(
Int (Internal)
Free running measurement
Man (Manual)
Triggers a measurement when
Ext (External) Triggers a measurement by external trigger signal input (through the external
trigger connector or handler interface).
To trigger a measurement in each mode, see \To Trigger a Measurement" later in this chapter.

To Set Trigger Delay Time

1. Press
.
2. Enter the desired trigger delay time using the numeric ENTRY keys. (For example, to set
0.5 s, press
3. Press
to set the value and to exit.
.
) points to the desired trigger mode.
9
.) You can set the trigger delay time from 0 s to 9.999 s.
is pressed.
Operating the 4339B
2-9