Checking Calibration Status; Execution Status Of Error Correction For Each Channel; Execution Status Of Error Correction For Each Trace - Agilent Technologies ENA Series User Manual

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Calibration

Checking Calibration Status

Checking Calibration Status

Execution status of error correction for each channel

You can check the execution status of error correction for each channel with the error
correction status.
The error correction status is indicated in the channel status bar in the lower part of the
window by the symbols in the below table.
Symbol
Execution status of error correction
(displayed in blue)
Error correction: On (enabled for all traces)
Cor
(displayed in gray)
Error correction: On (enabled for some traces)
Cor
(displayed in gray)
Error correction: Off
Off
(displayed in gray)
Error correction: On (no calibration data)
---
(displayed in blue)
Error correction: On (Interpolation is being executed or the IF
C?
bandwidth, power level, power range, sweep time, sweep delay time,
sweep mode, or sweep type is different from that when the calibration
was executed.)
(displayed in blue)
Error correction: On (Extrapolation is being executed.)
C!

Execution status of error correction for each trace

You can check the status of the error correction actually executed for each trace with the
trace status area.
For a trace for which error correction is executed, the applied calibration type is indicated
in the trace status area by the symbols in the table below.
Symbol
Calibration type
Open response calibration
RO
Short response calibration
RS
Thru response calibration
RT
Enhanced response calibration
ER
1-port calibration
F1
Full 2-port calibration/2-port TRL calibration
F2
F3
Full 3-port calibration/3-port TRL calibration
F4
Full 4-port calibration/4-port TRL calibration
*1.The simplified full 3/4-port calibration and the simplified 3/4-port TRL calibration (that
acquire the calibration coefficients skipping part of measurement) are not discriminated
from the normal full 3/4-port calibration and 3/4-port TRL calibration, and
displayed.
If none of the symbols described above is displayed, error correction is not executed for the
102
*1
*1
or
is
F3
F4