25
Measure Commands

:MEASure:WINDow

Command
:MEASure:WINDow {ZOOM | {MAIN | ALL}}, {MEASN}
The :MEASure:WINDow command specifies whether measurements are
made in the ZOOM window (measurement gating) or over the entire
acquisition (MAIN or ALL). The MAIN and ALL parameters are equivalent.
Example
This example gates Measurement 1 to the zoom window.
myScope.WriteString ":MEASURE:WINDow ZOOM, MEAS1"
Query
:MEASure:WINDow? {MEASN}
This query returns whether the measurement is being performed on the
zoomed portion of the waveform (ZOOM) or the entire acquisition (MAIN
or ALL).
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference