:MEASure:STATistics:INCRement
Command Syntax
:MEASure:STATistics:INCRement
This command updates the statistics once (incrementing the count by one)
using the current measurement values. It corresponds to the front panel
Increment Statistics softkey in the Measurement Statistics Menu. This
command lets you, for example, gather statistics over multiple pulses
captured in a single acquisition. To do this, change the horizontal position
and enter the command for each new pulse that is measured.
This command is only allowed when the oscilloscope is stopped and quick
measurements are on.
The command is allowed in segmented acquisition mode even though the
corresponding front panel softkey is not available.
See Also
•
•
•
•
300
(see
page
658)
"Introduction to :MEASure Commands"
":MEASure:STATistics"
":MEASure:STATistics:RESet"
":MEASure:RESults"