Model 4200A-SCS Parameter Analyzer Reference Manual
LPT commands for general operations
General operation commands include commands to control timing, execution, communications, and
test status.
clrscn
This command clears the measurement scan tables associated with a sweep.
Usage
int clrscn(void);
Details
When a single sweepX command is used in a test sequence, there is no need to program a clrscn
command because the execut command clears the table. The clrscn command is only required
when multiple sweeps and multiple sweep measurements are used in a single test sequence.
Example
double res1[14], res2[14];
.
conpin(SMU1, 1, 0);
conpin(SMU2, 2, 0);
conpin(GND, 3, 0);
forcev(SMU1, 4.0); /* Apply 4 V to gate. */
smeasi(SMU2, res1); /* Measure drain current in */
/* each step; store results */
/* in res1 array. */
sweepv(SMU2, 0.0, 14.0, 13, 2.0E-2); /* Make */
/* 14 measurements */
/* over a range of 0 V to 14 V. */
clrscn(); /* Clear smeasi. */
forcev(SMU1, 5.0); /* Apply 5 V to gate. */
smeasi(SMU2, res2); /* Measure drain current in */
/* each step; store results in */
/* res2 array. */
sweepv(SMU2, 0.0, 14.0, 13, 2.0E-2); /* Perform */
/*14 measurements */
/* over a range 0 V through 14 V. */