Model 4200A-SCS Parameter Analyzer Reference Manual
Status-code bit map for pulse_fetch
Bit
31
30
29
28
27 - 24
23 - 20
19 - 16
15 - 12
11 - 10
9 - 8
7 - 4
3 - 2
1 - 0
4200A-901-01 Rev. C / February 2017
Summary or description
Reserved
Sweep skipped
Load-line effect compensation (LLEC)
enabled (only valid when LLEC is
enabled)
LLEC status
RPM mode settings
Reserved
Measurement type
Current threshold, voltage threshold,
power threshold, and source compliance
Current measure overflow
Voltage measure overflow
Current measure range
Voltage measure range
Channel number
Section 13: LPT library function reference
Value (bit pattern)
Reserved bit for future use
0 = Not skipped
1 = Skipped
0 = Failed
1 = Successful
0 = Disabled
1 = Enabled
0 (0000) = No RPM
1 (0001) = RPM
2 (0010) = Bypass; PMU
3 (0011) = Bypass; SMU
4 (0100) = Bypass; CVU
All other values (bit patterns) reserved
Reserved bits for future use
1 (0001) = Spot mean
2 (0010) = Waveform
All other values (bit patterns) reserved
0 (0000) = None
1 (0001) = Source compliance
2 (0010) = Current threshold reached or surpassed
4 (0100) = Voltage threshold reached or surpassed
8 (1000) = Power threshold reached or surpassed
0 (00) = No overflow
1 (01) = Negative overflow
2 (10) = Positive overflow
0 (00) = No overflow
1 (01) = Negative overflow
2 (10) = Positive overflow
0 (0000) = 100 nA (RPM only)
1 (0001) = 1 µA (RPM only)
2 (0010) = 10 µA (RPM only)
3 (0011) = 100 µA
4 (0100) = 1 mA (RPM only)
5 (0101) = 10 mA
6 (0110) = 200 mA
7 (0111) = 800 mA
All other values (bit patterns) reserved
0 (00) = 10 V
1 (01) = 40 V
1 (01) = Ch1
2 (10) = Ch2
Value 0 (00) not used