System Mode SMU default settings

When KXCI mode is started, the SMUs are in a default state that mimics the Keysight Model 4145B,
as shown in the channel definition and source setup tables below. This means that each SMU is
active and part of the test, whether or not the test used all the SMUs in the 4200A-SCS chassis.
This may be undesirable, as many tests use a different number of SMUs or SMUs in a different state
from the Keysight 4145 default. To avoid the complication of building a KXCI test starting with the
default setup, use the following commands to remove the SMUs from the setup:
DE CH1 CH2 CH3 CH4 VM1 VM2 VS1 VS2
The above shows the method for a system with eight SMUs. A system with only four SMUs would not
require the last four definitions:
DE CH1 CH2 CH3 CH4
Channel Definition (Page DE) KXCI defaults
Instrument
SMU1
SMU2
SMU3*
SMU4*
SMU5*
SMU6*
SMU7*
SMU8*
* If there are less than eight SMUs, only the SMUs in the chassis
are programmed to the defaults shown.
Source Setup (Page SS) KXCI defaults
Name
Sweep mode
Start
Stop
Step
Number of steps 101
Compliance
Constant
V1 COM
V4 V
VS1 V
VS2 V
4200A-901-01 Rev. C / February 2017
Name
Source
V
I
Mode
Function
V1
I1
COM
CONSTant
V2
I2
I
VAR2
V3
I3
V
VAR1
V4
I4
V
CONSTant
VS1
N/A
V
CONSTant
VS2
N/A
V
CONSTant
VM1
N/A
N/A
N/A
VM2
N/A
N/A
N/A
VAR1
VAR2
(Command VR)
(Command IP)
V3
I2
Linear
Linear
0.00 V
20.00 µA
1.00 V
N/A
0.010 V
20.00 µA
5
100.0 mA
2 V
Source
Compliance
0.00 V
105.0 mA
0.00 V
100.0 mA
0.00 V
N/A
0.00 V
N/A
Section 9: Keithley External Control Interface (KXCI)
Command
CH
CH
CH
CH
VS
VS
VM
VM
Command
N/A
VC
SC
SC
9-51