Keithley 4200A-SCS Reference Manual page 797

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Section 9: Keithley External Control Interface (KXCI)
instrumentinfo
This command retrieves all information on the instrument cards in the 4200A-SCS.
Usage
int instrumentinfo(char *result, int maxlen, int *len);
result
maxlen
len
Details
This function returns all of the instrument-level information for the cards in the 4200A-SCS. The result
string contains the following information for each instrument card in the 4200A-SCS chassis:
Slot number
Instrument card ID
Model number
Serial number
Hardware version
Firmware version
Calibration date
Calibration due date
All of the information is comma-separated.
Example
Output for a 4200A-SCS with 3 instrument cards installed in slot 3 (4220-PGU), slot 5 (4225-PMU), slot 7
(4210-CVU):
slotno:3,name:VPU1,model:KIVPU4220,serialno:1254281,hwver:1.0,fwver:2.04,calda
te:Dec 19, 2009,caldue:Dec 19, 2016,
slotno:5,name:PMU1,model:KIPMU4225,serialno:1276563,hwver:1.0:544911,fwver:2.0
4,caldate:Nov 28, 2011,caldue:Nov 27, 2016,
slotno:7,name:CVU1,model:KICVU4210,serialno:Z005712,hwver:5.0:493083,fwver:2.1
2,caldate:Aug 09, 2010,caldue:Aug 09, 2016
Also see
None
9-86
String of results for the instruments
The maximum number of bytes that the result can stored in the buffer
Number of bytes returned by the function
Model 4200A-SCS Parameter Analyzer Reference Manual
4200A-901-01 Rev. C / February 2017

Advertisement

Table of Contents
loading

Table of Contents