Appendix K: Using a Signatone CM500 Prober

Probesubsites Clarius project example

The following procedure configures a Clarius project to execute testing and automatic wafer stepping
to all programmed subsites.
Use the 4200A-SCS to do this example. Use Clarius to open and run the probesubsites project
using the new configuration file, which will allow you to execute the project for this prober.
This project uses a Series 700 Switching System and the connect actions to change the instruments
connected to each pin without changing the physical configuration.
From Clarius:
1. Choose Select.
2. Select the Projects library.
3. Create the probesubsites project.
4. Select Configure.
5. In the project tree, select the prober-init action.
6. Set the mode to 6.
7. Set the subprobtype. If the CM500 prober is not at its first site, set the subprobtype to 1;
otherwise, set it to 0.
8. In the project tree, select probesubsites_1.
9. Select Run to execute the project.
K-18
Model 4200A-SCS Parameter Analyzer Reference Manual
Figure 826: Set prober-init mode parameters
4200A-901-01 Rev. C / February 2017