Keithley 4200A-SCS Reference Manual page 744

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Model 4200A-SCS Parameter Analyzer Reference Manual
YA
This command configures the Y1-axis of the graph.
Usage
YA 'AAAAAA', B, ±CCCC.CCC, ±DDDD.DDD
AAAAAA
B
CCCC.CCC
DDDD.DDD
Example
YA 'I1', 1, -5E-9, 5E-9
This command string:
Specifies that values from SMU channel I1 are to be plotted on the Y1-axis.
Sets up the Y1-axis to be scaled linearly between -5 nA and +5 nA.
Also see
Channel definition page (DE)
4200A-901-01 Rev. C / February 2017
The SMU channel name for the Y1-axis, up to 6 characters; must be one of the
SMU channel names that you specify on the channel definition (DE) page
Y1-axis scale type:
Linear scale: 1
Log scale: 2
Log scale absolute value: 3
Y1-axis minimum value:
Volts: ±9999
Amps: ±999
Y1-axis maximum value:
Volts: ±9999
Amps: ±999
(on page 9-13)