Section 6: Clarius
PMU: Allow unsettled measurements
When this option is selected, all 4225-PMU instrument cards ignore the minimum timing versus
measure range relationship. This is only recommended for advanced users, as spot mean results will
not be settled, which may cause a variety of operational issues, such as:
Inconsistent current measure range-changing on the 4225-PMU or 4225-RPM.
Lack of proper load-line effect compensation (LLEC) for the PMU.
Lack of correlation between PMU or PMU and RPM results and SMU results.
See
PMU minimum settling times versus current measure range
information.
Allow access to UTM UI editor
Select this option to access the user interface editor for user test modules (UTMs).
Refer to
Define the user interface for a user test module
editor.
To prevent unintentional changes to the user interface, it is good practice to disable the editor after
changes are complete.
Track site during stress test
Select this option to track sites during the stress test.
When this option is selected, during the subsite stress test:
The site identification box in the project tree is automatically updated to the presently started site.
For example, if the run site was set to 3, then after Run is selected, it is automatically reset to 1. It
changes to 2 as soon as testing switches to site 2.
All open subsite views are automatically switched to the presently running site. The Configure
and Analyze panes will match the presently running site.
If the subsite is opened, that also automatically converts to the present site during execution.
You can still change to other sites manually during execution.
My Projects Directory
Use this option to change the directory where your projects are stored. To create a new folder, right-
click and select New Folder.
Changing the project directory does not affect the open project or previously created projects. If you
save the open project, it is saved to its original location.
Existing projects remain in the directory in which they were previously stored. The default location is
C:\s4200\kiuser\projects.
You can move projects to the new location using Windows Explorer. When moving projects, move the
project folder and all its contents.
6-302
Model 4200A-SCS Parameter Analyzer Reference Manual
(on page 5-41) for additional
(on page 6-109) for detail on how to use the
4200A-901-01 Rev. C / February 2017