Keithley 4200A-SCS Reference Manual page 1006

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Model 4200A-SCS Parameter Analyzer Reference Manual
pulse_width
This command sets the pulse width for pulse output.
Usage
int pulse_width(int instr_id, long chan, double width);
instr_id
chan
width
Pulse modes
Standard
Details
Use the pulse_source_timing command to set the pulse width for the 4220-PGU and 4225-
PMU.
You can set the pulse width independently for each pulse card channel. The pulse_range
command is used to set pulse speed.
Pulse card pulse width is based on the full width at half-maximum method (FWHM). As shown below,
the pulse width is measured at the median (50 percent amplitude) point from the rising edge of the
pulse to the falling edge of the pulse.
4200A-901-01 Rev. C / February 2017
The instrument identification code of the pulse card, such as VPU1 or VPU2
Channel number of the pulse card: 1 or 2
Pulse width in seconds:
Fast speed (5 V): 10e-9 to (Period − 10e-9)
Slow speed (20 V): 250e-9 to (Period − 10e-9)
Default: 500e-9
Section 13: LPT library function reference