Keithley 4200A-SCS Reference Manual page 403

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Section 6: Clarius
Model 4200A-SCS Parameter Analyzer Reference Manual
Starting with default waveforms
The SegARBConfig dialog box can supply default waveforms. Select the Defaults button shown in
"Channel 2 SegARBConfig parameter configuration" in
Assigning SegARBConfig to a group
(on page
6-132).
You can use the created defaults to get started with a new UTM without having to input any
parameter values. Figure "Segment Arb Defaults Configuration" below shows the created blank
Segment Arb Default dialog box. To get a basic default waveform, select the Load Simple Pulse
button. This results in the four segment pulse shown in "UTM UI Editor Segment Arb Defaults
Configuration (after pressing Load Simple Pulse)" below.
The stop voltage of a segment must equal the start voltage of the next segment.
Figure 295: Segment Arb Defaults Configuration
6-134
4200A-901-01 Rev. C / February 2017