Section 9: Keithley External Control Interface (KXCI)
NR
This command sets the number of readings that can be made for time domain measurements.
Usage
NR AAAA
AAAA
Details
For time domain measurements, you can set the number of sample measurements that can be made.
The readings are stored in the buffer.
Example
NR 200
This command string sets up the 4200A-SCS to make 200 sample measurements.
Also see
None
DM
This command selects the Keysight 4145B display mode.
Usage
DMA
A
Details
The 4200A-SCS supports the Keysight 4145B graphics display mode and the Keysight 4145B list
display mode command.
The 4200A-SCS does not accept the matrix mode and schmoo mode commands (DM3 and DM4).
Example
DM1
This command string prepares the 4200A-SCS to receive graphics commands.
Also see
None
9-30
Number of measurements to make:
4200A command set: 1 to 4096
4145 Emulation command set: 1 to 1024
Display modes:
Graphics display mode: 1
List display mode: 2
Model 4200A-SCS Parameter Analyzer Reference Manual
4200A-901-01 Rev. C / February 2017