Ctsweep82 User Module - Keithley 4200A-SCS Reference Manual

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Model 4200A-SCS Parameter Analyzer Reference Manual

CtSweep82 user module

The CtSweep82 user module measures capacitance as a function of time at a certain bias.
Usage
status = CtSweep82(int Frequency, double Default_Bias, double Stress_Time, double
Test_Bias, double Sample_Time, int Reading_rate, int Num_Points, int Range590,
int Model590, int Filter590, char *CabComFile, int OffsetCorrect, char
*instr_id, int InputPin, int OutPin, double *CHF, int CHF_ArrSize, double
*G_or_R, int G_or_R_ArrSize, double *Time, int Time_ArrSize);
status
Frequency
Default_Bias
Stress_Time
Test_Bias
Sample_Time
Reading_rate
Num_Points
Range590
Model590
Filter590
CabCompFile
OffsetCorrect
instr_id
InputPin
OutPin
CHF
CHF_ArrSize
G_or_R
G_or_R_ArrSize
Time
Time_ArrSize
4200A-901-01 Rev. C / February 2017
Returned values are placed in the Analyze sheet; see Details
The measurement frequency:
100 kHz: 0
1 MHz: 1
DC bias applied before and after a C-t sweep (−20 V to +20 V)
Duration of the default bias before test bias is applied (0.001 s to 65 s)
Voltage bias for capacitance measurements (−20 V to +20 V)
Time delay between each sampling measurement (0.001 s to 65 s)
The reading rate used to acquire the measurements (1 to 4; see Details)
Number of sampling points (1 to 1350)
The measurement range for the 590 (1 to 4; see Details for valid range values)
The measurement model to use for high frequency measurement:
Parallel mode: 0
Series model: 1
Enable or disable the analog filter; see Details:
Disable the filter: 0
Enable the filter: 1
The complete name and path for the cable compensation file; see Details
Enable or disable an offset correction measurement:
Disable offset correction: 0
Enable offset correction: 1
KCon instrument ID; default is CMTR1; can be CMTR1 to CMTR4, depending on your
system configuration
The DUT pin to which the 5951 input terminal is attached (−1 to 72); if a value of
less than 1 is specified, no switch matrix connections are made; see Details
The DUT pin to which the 5951 output terminal is attached (−1 to 72); if a value of
less than 1 is specified, no switch matrix connections are made; see Details
Output; the measured array of high frequency capacitance values
Set to 1350
Output; the array of measured conductance (G) or resistance (R) values
Set to 1350
Output; the array of time from the 595 output for each measurement step
Set to 1350
Appendix D: Using a Model 82 C-V System
D-33