Section 13: LPT library function reference

dev_abort

This command programmatically ends a test from within the user module (aborts a test) that was started with the
pulse_exec command.
Usage
int dev_abort(NULL);
Pulsers
4220-PGU
4225-PMU
Pulse mode
Standard and Segment Arb
Details
Use this command to abort a pulse test from in a user module. This command is useful during a
longer pulse_exec test. Note that pulse_exec is nonblocking, so it is possible to fetch data during
a longer test. An example of this is a long stress/measure test, using seg_arb_sequence and
seg_arb_waveform, that evaluates degradation data intra-test. Evaluating this data from within the
user module may determine that a test should end. For example, if the degradation is greater than ten
percent (>10%), then end the test to save test time.
13-92
Model 4200A-SCS Parameter Analyzer Reference Manual
4200A-901-01 Rev. C / February 2017