Appendix D: Using a Model 82 C-V System
Procedure
For each range and test frequency specified by the input parameters:
1. You are prompted to open the circuit so that an offset capacitance measurement can be made.
2. Once the offset capacitance measurement is completed, you are prompted to connect the low
value capacitor for the selected range. The system performs the low capacitor compensation.
3. You are prompted to connect the high value capacitor for the selected range. The system does
the high value capacitor compensation.
4. You are prompted to reconnect the low capacitor.
5. The nominal and measured values are displayed in a dialog box.
6. Verify the values. If you are unsatisfied with the measurement, select Cancel to abort the
procedure. If you select Cancel, the cable compensation file is not affected.
7. When all selected ranges and frequencies have been compensated successfully, the cable
compensation values are saved.