Hp4284Ulib User Library - Keithley 4200A-SCS Reference Manual

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Appendix C: Using a Keysight 4284/4980A LCR Meter
A configured sweep is shown in the following figure.
A typical graph that is generated by this test is shown below.

HP4284ulib user library

You use the user modules in the HP4284ulib user library to control the Keysight 4284A or 4980A
LCR Meter. These user modules are summarized in the following table.
HP4284ulib user library
User module
Cmeas4284
(on page C-11)
CvSweep4284
C-8
Figure 577: C-V linear staircase sweep
Figure 578: Typical C-V curve for a MOS capacitor
Description
Makes a single capacitance measurement.
(on page C-9)
Makes capacitance versus voltage measurements using a staircase
sweep.
Model 4200A-SCS Parameter Analyzer Reference Manual
4200A-901-01 Rev. C / February 2017

Advertisement

Table of Contents
loading