Typical Preamplifier Matrix Card Connections - Keithley 4200A-SCS Reference Manual

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Model 4200A-SCS Parameter Analyzer Reference Manual
Appendix A: Using switch matrices

Typical preamplifier matrix card connections

The following figure shows typical preamplifier matrix card connections using local sensing. This
configuration is similar to the SMU configuration shown in the previous figure, except that
preamplifiers are added for low-current source-measure capabilities. The preamplifier FORCE
terminals are connected to the matrix card rows, while the DUT HI terminals are connected to the
matrix card columns. All 12 DUT LO terminals are connected together, and the common DUT LO
signal is connected to the ground unit FORCE terminal. Any preamplifier FORCE terminal can be
connected to any DUT HI terminal by closing the appropriate matrix crosspoint.
4200A-901-01 Rev. C / February 2017
A-9

Advertisement