Calling Kult User Libraries Remotely - Keithley 4200A-SCS Reference Manual

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Model 4200A-SCS Parameter Analyzer Reference Manual

Calling KULT user libraries remotely

KXCI contains a set of commands to call user libraries built by KULT on the 4200A-SCS from a
remote interface. Refer to
The commands for calling user modules are:
UL: usrlib
EX: execute
GN: get parameter
or output parameter. The parameter is specified by name.
GP: get parameter (by number)
specified input or output parameter. The parameter is specified by number. The first parameter
after the command is number one.
GD – get description
specified function. The function is specified by User Library and User Module.
UL
This command switches KXCI operation to the usrlib mode.
Usage
UL
Details
This command needs to be sent only once before any of the other commands to call user modules.
To switch back to normal KXCI command modes, send DE or US.
Send the UL command through GPIB or ethernet to change to the remote usrlib command set.
Also see
None
AB
This command aborts the execution of a module of the user library.
Usage
AB
Also see
EX
(on page 9-82)
4200A-901-01 Rev. C / February 2017
Keithley User Library Tool
(on page 9-81): Mode Switch: Switches KXCI to the usrlib mode.
(on page 9-82): Execute: Executes the specified module in a KULT user library.
(on page 9-83): Returns the parameter value (or values) for the specified input
(on page 9-84): Returns the parameter value (or values) for the
(on page 9-85): Get Parameter Description: Returns the description of the
Section 9: Keithley External Control Interface (KXCI)
(on page 8-1) for details on using KULT.
9-81