Section 13: LPT library function reference
The 4225-PMU or 4220-PGU returns to the following states:
•
The pulse mode is maintained. For example, if the pulse card is in Segment Arb mode, it will still
be in Segment Arb mode after the devint process is complete.
•
5 V and 10 mA ranges
•
If in pulse mode:
Period of 1.0 µs
Transition Times (Rise and Fall) of 100 ns
Width of 500 ns
Voltage high and low of 0 V
Load of 50 Ω
•
If in segmented arb mode, Start Voltage is 0 V
•
If in arbitrary waveform mode, Table Length is 100
Also see
clrcon
(on page 13-199)
clrscn
(on page 13-9)
clrtrg
(on page 13-11)
dev_abort
(on page 13-92)
devclr
(on page 13-67)
kibdefint
(on page 13-24)
13-206
Model 4200A-SCS Parameter Analyzer Reference Manual