Section 7: Keithley Configuration Utility (KCon)
Step 2. Add the equipment
To add equipment to the system configuration:
1. At the bottom of the Configuration Navigator, select Add External Instrument.
2. Select the specific instrument. An example of the selection of the System 82 Simultaneous CV
Capacitance Meter is shown in the figure below.
3. Select OK to add the instrument to the Configuration Navigator.
For descriptions of the options for your equipment, refer to
page 7-14).
Step 3. Set GPIB addresses
The GPIB address setting in KCon must match the actual GPIB address of the instruments in the
system. See the documentation for the instrument to determine the GPIB Address.
The address for some instruments is briefly displayed during the power-on sequence.
To set the GPIB address for an instrument:
1. In the Configuration Navigator, select the instrument.
2. Select the GPIB Address from the list. Addresses that are in use are displayed with asterisks (*)
next to them. The range of addresses is 0 to 30 (GPIB address 31 is reserved as the 4200A-SCS
controller address). If the selected GPIB address conflicts with the GPIB address of another
system component, a red exclamation-point symbol (!) is displayed next to the selected address.
You can programmatically read the GPIB address and other instrument properties from the system
configuration using the LPT library getinstattr function. Proper use of getinstattr allows you
to develop user libraries that are independent of the configuration. For more information, refer to
Keithley User Library Tool (KULT)
7-8
Model 4200A-SCS Parameter Analyzer Reference Manual
Figure 412: Add external equipment
(on page 8-1).
Instrument properties and connections
4200A-901-01 Rev. C / February 2017