Section 9: Keithley External Control Interface (KXCI)
:CVU:DATA:STATUS?
This command queries the status of the CVU when a test is complete.
Usage
:CVU:DATA:STATUS?
Also see
None
:CVU:DATA:TSTAMP?
This command queries the timestamp of the measurements of the CVU once a test is complete.
Usage
:CVU:DATA:TSTAMP?
Also see
None
Modeless commands
:CVU:CHANNEL chan
This command selects the CVU card on which subsequent CVU commands will act.
Usage
:CVU:CHANNEL chan
chan
Details
The majority of systems only have one CVU card. You can use this command to configure multiple
cards.
Also see
None
:CVU:MODEL
This command sets the measurement model for the selected CVU card.
Usage
:CVU:MODEL model
model
Also see
None
9-72
CVU card on which KXCI CVU commands act (default 1)
The model:
•
0: Z, theta
•
1: R + jx
•
2: Cp, Gp
•
3: Cs, Rs
•
4: Cp, D
•
5: Cs, D
Model 4200A-SCS Parameter Analyzer Reference Manual
4200A-901-01 Rev. C / February 2017