Section 9: Keithley External Control Interface (KXCI)
VM
This command defines channels that are used as voltmeters.
Usage
VMA, 'BBBBBB'
A
BBBBBB
Details
KXCI allows up to eight source-measure units (SMUs) to function solely as voltmeters. You can use
any channel for any voltmeter function between VM1 and VM9. For example, in a system containing
four SMUs, you can use SMU3 as VM7.
If you do not define one of the 4200A-SCS SMUs to emulate a VM, attempts to measure voltages
through the nonexistent VM result in data values of 9.000e+37.
If nothing is specified after the prefix and channel number, the channel is not used.
To measure voltage using the 4145B VM1...VMn function, define one of the 4200A-SCS SMUs to
emulate VM.
Example
VM1, 'VM1'
This command string defines the channel used for VM1 (specifies the name VM1).
Also see
KXCI Settings tab
Source setup page (SS)
Use the command strings for the SS page for the following operations:
VAR1 setup
VAR2 setup
VAR1' setup
List sweep setup
Constant SMUs setup
Constant VS setup
Set sweep hold time
Set sweep delay time
In order to send the following command strings to the 4200A-SCS, you must first select the source
setup page by sending the command:
SS
9-16
Voltmeter channel: 1 to 9; the assigned value for a voltmeter depends on how
instruments are mapped in KCon
User-specified name (up to 6 characters)
(on page 7-16)
Model 4200A-SCS Parameter Analyzer Reference Manual
4200A-901-01 Rev. C / February 2017