Model 4200A-SCS Parameter Analyzer Reference Manual

Probesubsites Clarius project example

The following is a step-by-step procedure to configure the manual prober so the probesubsites
Clarius project executes successfully. The user is responsible for the probe station set-up.

Use KCon to add a prober

On the 4200A-SCS, use KCon to add the prober to the configuration:
1. Open KCon.
2. At the bottom of the Configuration Navigator, select Add External Instrument. The Add External
Instrument dialog box is displayed.
4200A-901-01 Rev. C / February 2017
Figure 731: Add a prober in KCon
Appendix I: Using a manual or fake prober
I-7