Update The 4200A-Cviv Configuration; Save; System Configuration Summary - Keithley 4200A-SCS Reference Manual

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Model 4200A-SCS Parameter Analyzer Reference Manual

Update the 4200A-CVIV configuration

For more information on adding or removing 4200A-CVIV connections, refer to the "Configure and
install the 4200A-CVIV" section of the Model 4200A-CVIV User's Manual.
To update CVIV configuration:
1. Turn off system power.
2. Disconnect the 4200A-CVIV from the 4200A-SCS.
3. Make connections to the 4200-CVIV.
4. Turn on system power.
5. Start KCon.
6. Select the Update Preamp, RPM, and CVIV Configuration option.
7. Select Save.

Save

Saves changes to the system configuration. If you do not save the configuration changes, KCon
returns to the last saved configuration when you exit.

System Configuration Summary

System Configuration Summary displays the system configuration. From this display, you can save
the configuration or print it.
The sections of the system configuration are:
System Information: Provides a table containing the 4200A-SCS serial number, the network or
system name, and pertinent software and platform version information.
Instrumentation: Provides a table of properties and settings for each instrument in the system
configuration.
Connections: Provides a table of system connections. The connection table provides guidance
when making connections between the instrumentation and an optional switch matrix and test
fixture or probe station.
Select Save Configuration As to store the file in html format. You can view the html file in a web
browser.
To print the configuration, select Print Configuration.
4200A-901-01 Rev. C / February 2017
Section 7: Keithley Configuration Utility (KCon)
7-11