Model 4200A-SCS Parameter Analyzer Reference Manual
2. Ensure that the CabCompFile field has the same file directory path that is used in
capacitance source values (SaveCableCompCaps82)
3. Set the other fields to 8.
4. Click Run. The calibration source values entered into the action are placed into its spreadsheet.
5. Select Analyze. The sheet displays the values. An example spreadsheet is shown here.
Compensate for cable capacitance (CableCompensate82)
To compensate for cable capacitance:
1. In the project tree, select cable-compensate or cablecomp.
2. Select Configure.
3. Ensure that the CabCompFile field of the parameter list has the same file directory path that is
used in
4. Enable or disable cable compensation: Use the frequency and range fields to either disable (0) or
enable (1) cable compensation for the test frequencies and ranges. The following figure shows
cable compensation enabled for all ranges and test frequencies.
4200A-901-01 Rev. C / February 2017
Figure 587: display-cap-file spreadsheet with capacitor source values
Enter and save capacitance source values
Figure 588: CableCompensate82 user module
Appendix D: Using a Model 82 C-V System
(on page D-10).
(on page D-10).
Enter and save
D-11