Section 6: Clarius
Figure 292: SegARBConfig uses SeqList, SeqStartSeg, and SeqStopSeq to store sequence
In this example for the user module PMU_SegArb_ExampleFull, several parameters are common
across the two channels in the test:
SegTime (SegTime)
SegTrig (SegTrigOut)
SegMeas (MeasEnabled)
SegMeasType (MeasType)
SegMeasStart (SegMeasStart)
SegMeasStop (SegMeasStop)
SeqList (SeqList)
SeqStartSeg (SeqStartSeg)
SeqStopSeg (SeqStopSeg)
Assigning SegARBConfig to a group
As with other Control Types, to appear in the UTM Key Parameters pane, the SegARBConfig must
be assigned to a group. If a user module has more than one Segment Arb
a SegARBConfig control for each channel in a separate group. Choose which channel is configured
by selecting the specific variable name.
The figures below show the SegARBConfig for user module channel 1 and channel 2. For this
example, selecting the variable named StartVCH1 configures the SegStartV for channel 1, where
selecting the variable named StartVCH2 configures the SegStartV for channel 2.
6-132
Model 4200A-SCS Parameter Analyzer Reference Manual
®
4200A-901-01 Rev. C / February 2017