Probesites Clarius Project Example - Keithley 4200A-SCS Reference Manual

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Model 4200A-SCS Parameter Analyzer Reference Manual
8. Click Set Reference Die in the Setup Options window to open the Set Reference dialog box.
9. Move the pins over the pads of the die to be set as the reference die.
10. Zero out the X/Y, Column, and Row (click Zero X,Y button and Zero C,R).
If you want the columns and rows to be something other than 0,0 (1,1 for instance), edit values in Set
Reference dialog box as needed before clicking Done.
11. Click Done.

Probesites Clarius project example

The following is a step-by-step procedure to configure the 8860 so the probesites Clarius project
executes successfully.
On the pcBridge computer:
1. Use the pcWafer program to probe a single subsite on multiple dies.
2. Start pcWafer by clicking the pcWfr button in the pcLaunch window. The pcWfr window appears.
4200A-901-01 Rev. C / February 2017
Appendix H: Using a Micromanipulator 8860 Prober
Figure 705: Set Reference Die button
Figure 706: Set Reference dialog box
Figure 707: pcWfr button
H-17