Table 14-1 General Bus Commands - Keithley SourceMeter 2400 User Manual

Sourcemeter 2400 series source measure unit
Hide thumbs Also See for SourceMeter 2400:
Table of Contents

Advertisement

®
2400 Series SourceMeter
User's Manual
General bus commands
General commands are those commands, such as DCL, that have the same general mean-
ing regardless of the instrument.
Table 14-1
General bus commands
Command
REN
Goes into remote when next addressed to listen.
IFC
Goes into talker and listener idle states.
LLO
LOCAL key locked out.
GTL
Cancel remote; restore SourceMeter front panel operation.
DCL
Returns all devices to known conditions.
SDC
Returns SourceMeter to known conditions.
GET
Initiates a trigger.
SPE, SPD
Serial polls the SourceMeter.
REN (remote enable)
The remote enable command is sent to the SourceMeter by the controller to set up the
instrument for remote operation. Generally, the instrument should be placed in the remote
mode before you attempt to program it over the bus. Setting REN true does not place the
instrument in the remote state. You must address the instrument to listen after setting REN
true before it goes into remote.
The SourceMeter must be in remote in order to use the following commands to trigger and
acquire readings:
:INITiate and then :FETCh?
:READ?
:MEASure?
IFC (interface clear)
The IFC command is sent by the controller to place the SourceMeter in the local, talker,
listener idle states. The unit responds to the IFC command by cancelling front panel TALK
or LSTN lights if the instrument was previously placed in one of these states.
Note that this command does not affect the status of the instrument. Settings, data, and
event registers are not changed.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Remote Operations
Table 14-1
lists the general bus commands.
Effect on SourceMeter
14-7