Keithley 4200A-SCS Reference Manual page 743

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Section 9: Keithley External Control Interface (KXCI)
XN
This command configures the X-axis of the graph to plot an electrical parameter.
Usage
XN 'AAAAAA', B, ±CCCC.CCC, ±DDDD.DDD
AAAAAA
B
CCCC.CCC
DDDD.DDD
Example
XN 'V1', 1, -5, 5
This command string:
Specifies that values from SMU channel V1 are to be plotted on the X-axis.
Sets up the X-axis to be scaled linearly between −5 V and +5 V.
Also see
Channel definition page (DE)
XT
This command configures the X-axis of the graph to plot time domain values in seconds.
Usage
XT AAAA.AA, BBBB.BB
AAAA.AA
BBBB.BB
Example
XT 0, 10
This command string:
Specifies that time domain values are to be plotted on the X-axis.
Sets up the X-axis to be scaled between 0 and 10 s.
Also see
None
9-32
The SMU channel name for the X-axis; up to 6 characters long; must be one of the
SMU channel names that you specify on the channel definition (DE) page
X-axis scale type
Linear scale: 1
Log scale: 2
X-axis minimum value
Volts: ±9999
Amps: ±999
X-axis maximum value
Volts: ±9999
Amps: ±999
(on page 9-13)
X-axis minimum time value (in seconds): 0.01 to 9999
X-axis maximum time value (in seconds): 0.01 to 9999
Model 4200A-SCS Parameter Analyzer Reference Manual
4200A-901-01 Rev. C / February 2017

Advertisement

Table of Contents
loading

Table of Contents