Setauto - Keithley 4200A-SCS Reference Manual

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Model 4200A-SCS Parameter Analyzer Reference Manual
The following examples show the proper command sequence for using rtfary:
Example 1: Valid command
sequence for voltage sweep
smeasz
smeast
rtfary
dsweepv
Example
Programming example #2
Also see
None

setauto

This command re-enables autoranging and cancels any previous rangeX command for the specified instrument.
Usage
int setauto(int instr_id);
instr_id
Details
When an instrument is returned to the autorange mode, it will remain in its present range for
measurement purposes. The source range will change immediately.
Due to the dual mode operation of the SMU (v versus i), setauto places both voltage and current
ranges in autorange mode.
Example
double icer1;
double idatvg[25];
.
.
rangei(SMU1, 2.0E-9); /* Select manual range. */
delay(200); /* Delay after range change. */
measi(SMU1, &icer1); /* Measure leakage. */
.
.
setauto(SMU1); /* Enable autorange mode. */
lorangei(SMU1, 2.0E-6); /* Select 2 uA as minimum range */
/* during autoranging. */
delay(200); /* Delay after range change. */
smeasi(SMU1, idatvg); /* Setup sweep measurement */
/* of IDS. */
sweepv(SMU2, 0.0, 2.5, 24, 0.002); /* Sweep gate from 0 V to 2.5 V. */
Also see
None
4200A-901-01 Rev. C / February 2017
Example 2: Valid command
sequence for frequency sweep
smeasz
rtfary
dsweepf
(on page 13-195) returns the array of force values for the voltage sweep.
The instrument identification code
Section 13: LPT library function reference
13-85