Advertisement

Appendix B: Using a Model 590 C-V Analyzer
Csize
V
Vsize
G_or_R
G_or_Rsize
T
Tsize
B-22
Set to a value that is equal to or greater than the G_or_Rsize or number of voltage
steps in the sweep, or is equal to
((LastBias − FirstBias) /StepV) + 1; when this function is called from a
Clarius, this parameter is fixed at 1350
The array of bias voltages used
Set to a value that is equal to or greater than the G_or_Rsize or number of voltage
steps in the sweep, or is equal to
((LastBias − FirstBias) /StepV) + 1; when this function is called from a
Clarius, this parameter is fixed at 1350
Output:
When the parallel measurement model (1) is selected, G_or_R is the
measured conductance
When the series measurement model (0) is selected, this is the measured
resistance
Set to a value that is equal to or greater than the G_or_Rsize or number of voltage
steps in the sweep, or is equal to
((LastBias − FirstBias) /StepV) + 1; when this function is called from a Clarius
UTM, this parameter is fixed at 1350
The array of time stamps for each measurement step
Set to a value that is equal to or greater than the G_or_Rsize or number of voltage
steps in the sweep, or is equal to
((LastBias − FirstBias) /StepV) + 1; when this function is called from a Clarius
UTM, this parameter is fixed at 1350
Model 4200A-SCS Parameter Analyzer Reference Manual
4200A-901-01 Rev. C / February 2017