Keithley 4200A-SCS Reference Manual page 785

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Section 9: Keithley External Control Interface (KXCI)
:CVU:SPEED
This command sets the measurement speed for the selected CVU card.
Usage
:CVU:SPEED speed
:CVU:SPEED 3, delay_factor, filter_factor, aperture
speed
delay_factor
filter_factor
aperture
Details
The parameters delay_factor, filter_factor, and aperture are only used if speed is set to
custom.
Also see
None
:CVU:ACV
This command sets the AC drive level for specified CVU card.
Usage
:CVU:ACV aclevel
aclevel
Details
If in user mode, this takes immediate effect.
If in system mode, the value is buffered for use in all sweeps except AC Voltage sweep (which is set
with the :CVU:SWEEP:ACV command).
Also see
:CVU:SWEEP:ACV
9-74
Applies one of four defined speed selections:
0: Fast
1: Normal
2: Quiet
3: Custom
0 to 100
0 to 100
0.006 to 10.002 PLCs
10 mV to 100 mV
(on page 9-68)
Model 4200A-SCS Parameter Analyzer Reference Manual
4200A-901-01 Rev. C / February 2017

Advertisement

Table of Contents
loading

Table of Contents