Appendix A: Using switch matrices
Model 4200A-SCS Parameter Analyzer Reference Manual
7174A connections for remote sensing
The next figure shows how to connect instrumentation for remote sense operation. Since there are
not enough matrix rows, the instruments are connected to the matrix columns. In this configuration,
two switch relays are closed to complete a path from an instrument to a device under test (DUT). With
five DUT matrix cards installed in a Series 700 Switching System mainframe, up to 30 DUT pin-pairs
can be used.
In the next figure, the
C-V Analyzer signal paths
(on page A-20) for the Keysight Model 4980A and
Keysight Model 8110A pulse generator signal path
(on page A-23) show how signals are routed
through 7174A matrix switches to a DUT.
Figure 524: Remote sense test system using 7174A matrix cards
A-6
4200A-901-01 Rev. C / February 2017