Appendix B: Using a Model 590 C-V Analyzer
C
Csize
G_or_R
G_or_Rsize
T
Tsize
Details
This user module performs a capacitance versus time (Ct) sweep. The figure below shows the default
parameters for the ctsweep-590 UTM, which uses the CtSweep590 user module.
B-18
Output: The measured array of capacitance values
Set to a value that at minimum is equal to the Count; if in doubt, set to 1350
Output:
When the parallel measurement model (1) is selected, G_or_R is the
measured conductance
When the series measurement model (0) is selected, this is the measured
resistance
Set to a value that at minimum is equal to the Count; if in doubt, set to 1350
The array of time stamps for each measurement step
Set to a value that at minimum is equal to the Count; if in doubt, set to 1350
Figure 562: Starting KULT