Model 4200A-SCS Parameter Analyzer Reference Manual
11. Enter the name v_sweep_chk for the UTM. You will use the v_sweep_chk UTM to execute the
VSweep user module.
12. Select Configure.
13. In the right pane, from the User Libraries list, select the my_2nd_lib library.
14. From the User Modules list, select the v_sweep_chk user module. A default schematic and
group of parameters are displayed for the UTM.
15. For Vforce, select Enter Values and enter the sweep values.
16. Select Run.
17. Select Analyze.
At the conclusion of execution, review the results in the Analyze worksheet. If you connected a 1 kΩ
resistor between SMU1 and GNDU, used the default UTM parameter values, and executed the UTM
successfully, the results should be similar to the results in the following figure.The current/voltage
ratio for each row of results should be approximately 1 mA / V.
4200A-901-01 Rev. C / February 2017
Figure 442: Checking the VSweep user module
Section 8: Keithley User Library Tool (KULT)
8-33