Remove An External Instrument; Validate Configuration - Keithley 4200A-SCS Reference Manual

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Model 4200A-SCS Parameter Analyzer Reference Manual
Step 4. Save configuration
To save the KCon configuration:
Select Save.
Step 5. Validate configuration
This step checks communications between the 4200A-SCS and the attached equipment.
To validate configuration:
1. Select Validate Configuration.
The verification report is displayed on the screen.

Remove an external instrument

To remove an external component from the system configuration:
1. Select the component.
2. Select Remove External Instrument. A confirmation dialog box is displayed.
3. Select OK.
You cannot remove internal instruments.

Validate Configuration

Validate Configuration tests the system configuration to determine if there are configuration conflicts
or communication problems between the instrumentation and the 4200A-SCS. This test can validate
most of the supported internal and external instruments. However, the following instruments are not
automatically verified when the configuration is validated:
Probe stations
Test fixtures
General purpose test instruments
When you run Validate Configuration, KCon communicates with the instruments to verify that the
physical configuration matches the KCon defined configuration. The report is displayed in the KCon
Work Area. If KCon detects configuration conflicts, it displays corresponding error messages in the
Work Area.
Clarius automatically validates the configuration when it starts up. If Clarius detects conflicts, it
displays an error message that instructs you to resolve the conflicts using KCon.
4200A-901-01 Rev. C / February 2017
Section 7: Keithley Configuration Utility (KCon)
7-9