Section 9: Keithley External Control Interface (KXCI)
:CVU:SWEEP:DCV
This command configures the CVU to sweep DC voltage and sample Z-measurements for the selected CVU card.
Usage
:CVU:SWEEP:DCV dcvstart, dcvstop, dcvstep
dcvstart
dcvstop
dcvstep
Details
The other parameters (AC voltage level, frequency, and so on) are set by their respective commands.
Valid voltage levels are -30 V to +30 V.
Also see
None
:CVU:SWEEP:ACV
This command configures the CVU to sweep AC voltage and sample Z-measurements for the selected CVU card.
Usage
:CVU:SWEEP:ACV acvstart, acvstop, acvstep
acvstart
acvstop
acvstep
Details
Valid AC voltage levels are 10 mV to 100 mV. The other parameters (DC voltage level, frequency,
and so on) are set by their respective commands.
Also see
None
:CVU:DELAY:STEP
This command sets the hold time for the CVU test on the selected card for the :CVU:SWEEP:FREQ and
:CVU:SWEEP:DCV operations.
Usage
:CVU:DELAY:STEP stepd
stepd
Also see
:CVU:SWEEP:DCV
:CVU:SWEEP:FREQ
9-68
Start voltage
Stop voltage
Voltage step size
Start AC voltage
Stop AC voltage
AC voltage step size
Hold time to apply the pre-soak value (set by the :CVU:SOAK:DCV command); valid
values range from 0 to 999 s
(on page 9-68)
(on page 9-67)
Model 4200A-SCS Parameter Analyzer Reference Manual
4200A-901-01 Rev. C / February 2017